18
Function
Selecting Mode
Function Name
Key Operation & WGA-670B
Movement in Function Setting Mode
Functional
Description
F-13 BV
Select
BV set value 0: 10 V
1: 2 V
Select either of the two and press the SET
key.
•A function that
selects bridge
exciter voltage.
•It is set to 1(2 V)
prior to
shipment.
F-14
TEDS ON/OFF
Indicates the following items (1) and (2) .
(1) Sensitivity registration with power ON
0: OFF
1: ON
(2) Digital ZERO operation with power ON
0: OFF
1: ON
Select and change numerics in items (1) and
(2) and press the SET key.
(Enabled only for WGA-670B-7.)
•A function
whether or not
sensitivity
registration or
digital Zero is
conducted by
reading TEDS
information with
the power ON.
• Factory default is
‘0-0.’
Summary of Contents for WGA-670B
Page 3: ......
Page 56: ...51 9 2 OUTSIDE DRAWING 1 With No Option WGA 670B 0 ...
Page 57: ...52 2 With BCD Option WGA 670B 1 ...