15
Function
Selecting Mode
Function Name
Key Operation & WGA-670B
Movement in Function Setting Mode
Functional
Description
F-07
Smoothing
Indicates the following items (1), (2), and
(3).
(1) Number of (2)Minimum Scale
Moving Averages
0 : None 0 : Displays every 1 scale.
1 : 8 times 1 : Displays every 2 scales.
2 : 16 times 2 : Displays every 5 scales.
3 : 32 times 3: Displays every 10 scales.
4 : 64 times 4 : Displays every 20 scales.
5 : 128 times 5 : Displays every 50 scales.
6 : 256 times 6 : Displays every 100 scales.
7 : 512 times 7 : Displays every 200 scales.
8 : 1024 times 8 : Displays every 500 scales.
9 : 2048 times 9 : Displays every 1000 scales.
(3) Analog filter
0: 100 Hz
1: 10 Hz
Select and change numerics in items (1), (2)
and (3) and press the SET key.
•A function that
stables the
indication by
smoothing
varying signals
issued from the
transducer.
•The smoothing
function is
classified into the
following 3
types.
Number of
moving averages:
The number of
times to average
the measured
value.
Minimum scale:
Minimum unit
that changes the
indicated value.
Analog filter:
Cuts the extra
high frequency
transferred
from the
transducer.
Factory default is
0: 100 Hz .
F-08
Additional Value Set the desired additional value and press
the SET key. The entered value is added to
the measured value and indicated.
●
If digital ZERO function is activated
with the additional value set to the
measured value, the determined value is
indicated on the indicator.
●
Setting range: ±19999
•A function that
adds the desired
numerical value
to the measured
value.