HM-JACKarc Operation Manual
Analysis Operations
5 - 6 Rev.4
Reagent bottle
5.2.4 Preparing the Reagent
Set the bottle of EXTEL HEMOAUTO
・
HS Latex (referred to below as
"Reagent") in the Main Unit.
See "3.2 Part Names" for the names of the parts.
Read the Instructions for Use provided with the Reagent before
handling it.
Do not use Reagent that has expired, or under improper storage
condition or that is mixed with foreign substances or contaminated
with other chemicals. Accurate analysis results may not be obtained
if the Reagent is not prepared correctly.
Do not mix the Reagent from bottle to bottle in any case.
Setting the Reagent in place
Handling Precautions: Prepare sufficient volume of Reagent for the analysis to be performed.
1.
Let the Reagent warm to room temperature, and then mix it well and
gently by repeated inversion.
Handling Precautions: Mix the Reagent by repeatedly inverting the bottle gently. If air bubbles
mix into the Reagent, accurate measurement results may not be
obtained.
2.
Uncap the Reagent bottle and set the bottle in the Reagent Bottle
Holder next to the injection assembly.
Caution
Bottle Holder
Reagent Bottle
Reagent Bottle