HM-JACKarc Operation Manual
Analysis Operations
5 - 27 Rev.4
3.
Touch the column of
Protect
. A numeric keypad will appear. Then input
“
9466
” and touch
ENTER
. Then the column will change into
Clear
.
■
Inputting data for Calibrator
1.
After clearing input protection, go to the System Menu.
2.
Touch
Item Parameter
, and then
Next Page
at the bottom of the Item
Parameter screens twice to display Item Parameter (3/3).
3.
On the screen, from No.1 STD Revise Conc1 as the top to No.2 STD
Revise Conc2 as the bottom, they are the parameters for the data for
Calibrators to be input.
・
No.1 STD Revise Conc1 : Conc1 for Calibration Curve No.1
・
No.1 STD Revise Conc2 : Conc2 for Calibration Curve No.1
・
No.2 STD Revise Conc1 : Conc1 for Calibration Curve No.2
・
No.2 STD Revise Conc2 : Conc2 for Calibration Curve No.2
Touch the corresponding column. Then input the value with the displayed
numeric keypad in accordance with the Calibrator Card.
4.
Go to Calibration screen and confirm that the displayed data are the same
as the ones in the Calibrator Card.
5.
After confirmation, touch
Clear
column in the User Parameter screen to
change it back to
Protect
to avoid any unexpected change.
■
Inputting I.S.T of Master Curve
1.
After clearing protection, touch
Calibration curve
icon on the Main Menu.