HM-JACKarc Operation Manual
Analysis Operations
5 - 23 Rev.4
5.4 Shut Down Procedure
Finishing each measurement, the Analyzer removes the used cell from
the Reaction Table to the Cell Waste Box, picks up a new cell from the
Cell Cassette and set to the empty position. Chimes can be heard for
ten seconds when all the measurement is completed.
5.4.1 Clearing up after finishing analysis
1.
Performing Wash.
In case the next start-up is planned more than one week later, replace
the Buffer bottle with a bottle filled with distilled water or equivalent
solution before performing this process. Touch
Main Menu
icon, then
touch
Wash
on the Main Menu.
2.
Putting away the Reagent and Samples.
Once analysis is complete, discard all the used Cells in the Cell Waste
Box and Samples on the Sample Rack Tray.
Remove the Reagent from the Main Unit, put the cap back and store
them in a chemical storage refrigerator.
Waste materials should be handled as medical waste and disposed
of by recognized laboratory procedures.
For the storage of the Reagent and Buffer, please comply with
instructions of the Instructions for Use provided with the products.
Caution