HM-JACKarc Operation Manual
Functions of Each Screen
4 - 8 Rev.4
Sample Status field
The status of each sample is displayed.
・
Sel.
:
(Not in use)
・
No.
:
Sample No.: 1-320
・
Pos.
:
Attribute or Position Number
STD1 : Calibrator 1
STD2 : Calibrator 2
CTLL : Control L
CTLH : Control H
Cup
: Sample Cup(
as optional
)
Samples : rrr-pp
rrr: Rack No. 001-999 (rack barcode)
pp: Position No. within rack No. 1-10
・
ID CODE
When Sample barcode is enabled, Sample barcode for each
sample is displayed. When Sample barcode is disabled, User
ID is displayed.
・
Av.
:
Attribute or repetition number is displayed.
Calibrator 1 and 2 = Average number of Calibration
(Set as an item parameter: 1-3)
Control L = Repetition number of Control L
(Set as an item parameter: 1-3)
Control H = Repetition number of Control H
(Set as an item parameter: 1-3)
Cup = Repetition number of Sample Cup(
as optional
)
(Set as an item parameter: 1-3)
Patient Samples (one-day method) = 1
Patient Samples (two-day method) = 1 (1st day), 2 (2nd day)
・
Data
:
Results of the Sample is displayed.
・
Alarm
:
If an event occurred for the Sample, a flag is displayed in this
column. Refer to Error List for details of the code.
■
Status and Operation Message
The status of the analyzer and supplementary message is displayed
timely in this area during operation.
Latex column
Touch this column to input the volume of Latex Reagent. Available test
with the Latex Vial set to the Instrument is displayed and counted down
accordingly. See
“Setting Reagent volume” on page 5 - 6 for details.
■
Analysis end time
Estimated remaining time of the run is displayed in this column. This
column is not in use after finishing the run.