HM-JACKarc Operation Manual
Analysis Operations
5 - 14 Rev.4
5.2.7 Registration of Master Curve and Calibration Card
Registration of Master Curve and Calibrator Card are required when
their lots are changed.
1.
Touch the
Calibration Curve
icon to display the Calibration Curve
screen. (See 4.1.6 Calibration for reference)
2.
Touch the
Barcode
button to display the Master Curve/Calibrator
Registration screen.
3.
With above screen displayed, scan the barcodes.
When the Reagent lot is changed, register the barcodes on the Master
Curve card provided with the EXTEL HEMO
・
AUTO HS Latex in order
from the top. The Reagent lot number, the expiration date, and the Master
Curve parameters from 1 to 7 are input sequentially.
When the Calibrator lot is changed, register the barcodes on the Calibrator
Card provided with the EXTEL HEMO
・
AUTO HS Calibrator in order from
the top. The Master Curve, input data for Calibrators, lot number and
expiration date are input. (Refer to 5.5.1 Reading Barcodes with Handheld
Scanner)
Master Curve Card
Calibrator Card