HM-JACKarc Operation Manual
Analysis Operations
5 - 21 Rev.4
5.3 Analysis
It is mandatory to ensure results of Control L and H are within specified
range provided by Instructions for Use of EXTEL HEMO
・
AUTO HS
Control before proceeding with patient sample analysis.
Before starting Analysis, check the following.
Consumables, Reagent and Buffer are sufficient for the analysis.
Racks are loaded correctly.
The Main Cover is closed.
1.
Starting analysis
Touch the
Start
icon on the Order screen to start analysis. Analysis
screen will be displayed on the screen.
2.
Reporting results
During analysis, the results will be printed out real time.
■
Adding samples during analysis
When all the sampling finished, chimes ring and the message
“Remove
I-Tray for Sample addition
” appears. Racks can be added (or replacing
Tray) followed by this message, and then press
Start
icon to resume the
analysis.
■
Next batch analysis
The screen below is an example after finishing a batch of measurement.
To start the next batch, set samples, touch
To Order
button to go back
to Order screen, and then touch
Start
.
Caution