1.
Check if the Control Transformer button is pulled
out.
4.
Is there 120 VAC between #1 & #2 on the 120 VAC Terminal Strip? If not, check
the fuse on the output side of the Control Transformer. If fuse is good, check power
coming out of Control Transformer.
5.
If no power on the output side, and there is power going into the top of the Control
Transformer, replace the Control Transformer.
6.
If there is power at the Control Transformer, check the wiring of the black and
white wire going from the Control Transformer to the 120 VAC Terminal Strip.
7.
If there is no power between #1 and #2, check the secondary side of the trans-
former.
a.Check between X1 and X2. If no power is measured it is a bad transformer.
b.If there is power at X1 and X2, check the other side of the fuse. If now
power, replace the fuse.
2.
Is the Disconnect Switch on the main electrical cabinet set to
ON?
Switch
3.
Is there 208, 220, 440, or 575 VAC to the top side (input) of the Control Trans-
former? If not, check the fuses at the Fuse Block, and the contacts on the Control
Transformer button on the switch panel.
Input High
Vac
Output
120 Vac
Typical Control Transformer
Control Power Light OFF
Troubleshooting with the Status Light Panel
4-24
990-HQ Service Manual
Summary of Contents for 990-HQ
Page 4: ...KVAL 990 HQ Service Manual ...
Page 9: ...Notes ...
Page 29: ...Safety Sign Off Sheet 1 20 990 HQ Service Manual ...
Page 38: ...Network System Overview 2 9 990 HQ Service Manual ...
Page 65: ...Lubrication Points on the 990 HQ 3 27 990 HQ Service Manual FIGURE 3 18 Bearings Back Section ...
Page 79: ...Collet Torque Values 3 41 990 HQ Service Manual ...
Page 80: ...Appendix Conversion Chart 3 42 990 HQ Service Manual Appendix Conversion Chart ...
Page 81: ...Appendix Conversion Chart 3 43 990 HQ Service Manual ...
Page 108: ...Troubleshooting with the Status Light Panel 4 27 990 HQ Service Manual ...
Page 111: ......