Zero calibration skip
When shipped from the factory, the “Calibration (Excluding Zero Cal.)” button is enabled on the <Calibration>
screen. If configuring to ensure that zero calibration is performed, disable this function.
Memo
Zero calculation skippable is set to “Disable (Perform zero calibration)” when the instrument is shipped from the
factory.
Notes
When using the MAV target mask (with glass), it is necessary to set “Zero cal. skippable” to “Disable” because
the amount of stray light tends to fluctuate.
Operating Procedure
Start the procedure from the <Setting> - <Calibration setup> screen.
1
Use [
▲
] or [
▼
] to move the cursor to “Zero cal.
skippable”, and then press the [Confirmation] key.
2
Use [
▲
] or [
▼
] to move the cursor to select
“Disable” or “Enable”.
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Calibration setup> screen
.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Calibration setup> screen.
Summary of Contents for CM-23d
Page 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Page 30: ...27...
Page 82: ...79...
Page 111: ...108 Setting 3 MEMO...
Page 112: ...109...
Page 132: ...129...
Page 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Page 138: ...135...
Page 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Page 142: ...139 Dimensions Unit mm...
Page 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...