INSTALLATION INSTRUCTIONS
2-29
2004 December Rev.S 1908110
4. Select
Modem/Connection Properties
and then enter the following settings:
a. Bits per second:
38,400
b. Parity:
None
c. Data bits:
8
d. Stop bits:
1
e. Flow control:
SOFTWARE
5. Click [OK].
6. Select the “Data” tab followed by the “Transfer Protocol” button, and then
RAW ASCII
for the “Current
Transfer Protocol”
7. Click [OK].
8. Energize the IMAGER and insert the film saver onto the Film Tray.
9. From the “Service Test” menu, check that the Service Port Baud Rate is set to match the terminal
program baud rate (e.g. 38.4 K BPS).
10. From the SERVICE COMPUTER, type:
@
to check the connection.
The Error Log from the IMAGER will then be displayed on the SERVICE COMPUTER.
11. Click on the “Send File” icon.
12. In the “Look-in” field, enter the drive and folder where the file is located.
13. In the “Files of Type:” field, select All Files (*.*).
14. In the “Filename:” field, enter the name of the file you want to send, and click [Open].
Note
The S-records should begin transferring and, after a short delay, the IMAGER will return a (.) for
each group of several records received successfully.
15. If X’s are received, stop the transfer.
Note
X’s indicate an error in the received record and the transfer will not be successful.
a. De-energize and energize the IMAGER.
Note
You do not have to wait until the self--test is complete before continuing. You may continue after
rec++ ATSO =1 from the IMAGER.
b. Restart the procedure at step 10.
Note
After a successful transfer, the IMAGER sends the following messages, then boots and does a self
test after it energizes:
Reprogramming IMAGER FEEPROM
Erasing Sector XxXXXXXXXX
Writing Data to Sector XxXXXXXXXX
Summary of Contents for dryview 8300
Page 1: ...SERVICE MANUAL Kodak DryView 8300 LASER IMAGER 1908110 78 6970 7393 6 12 04 Rev S ...
Page 6: ...SERVICE MANUAL iv 1908110 2001 March Rev M BLANK PAGE ...
Page 8: ...SERVICE MANUAL vi 1908110 2001 March Rev M BLANK PAGE ...
Page 48: ...SERVICE MANUAL 2 30 1908110 2004 December Rev S Reprogramming Operation Complete ATSO 1 ...
Page 82: ...SERVICE MANUAL 3 20 1908110 2004 December Rev S BLANK PAGE ...
Page 108: ...SERVICE MANUAL 4 26 1908110 2004 March Rev R Figure 4 31 Figure 4 32 ...
Page 122: ...SERVICE MANUAL 4 40 1908110 2004 March Rev R Figure 4 50 ...
Page 150: ...SERVICE MANUAL 5 24 1908110 2004 December Rev S BLANK PAGE ...
Page 153: ...THEORY 6 3 2002 March Rev P 1908110 Figure 6 3 Heater Control Process ...
Page 154: ...SERVICE MANUAL 6 4 1908110 2002 March Rev P Figure 6 4 Calibration Process ...
Page 155: ...THEORY 6 5 2002 March Rev P 1908110 Figure 6 5 Main Print Control Routine Sheet 1 of 2 ...
Page 157: ...THEORY 6 7 2002 March Rev P 1908110 Figure 6 6 Wait for OK to Feed ...
Page 158: ...SERVICE MANUAL 6 8 1908110 2002 March Rev P Yes Figure 6 7 Is Film Path Clear ...
Page 159: ...THEORY 6 9 2002 March Rev P 1908110 Figure 6 8 Is There a Page to be Printed ...
Page 160: ...SERVICE MANUAL 6 10 1908110 2002 March Rev P Figure 6 9 Feed Film ...
Page 161: ...THEORY 6 11 2002 March Rev P 1908110 Figure 6 10 Feed Film to Separators ...
Page 162: ...SERVICE MANUAL 6 12 1908110 2002 March Rev P Figure 6 11 Start Feed ...
Page 163: ...THEORY 6 13 2002 March Rev P 1908110 Figure 6 12 Pull Back Lower Sheets ...
Page 164: ...SERVICE MANUAL 6 14 1908110 2002 March Rev P Figure 6 13 Pull All Sheets Back into Film Tray ...
Page 165: ...THEORY 6 15 2002 March Rev P 1908110 Figure 6 14 Check for Multifeed ...
Page 166: ...SERVICE MANUAL 6 16 1908110 2002 March Rev P Figure 6 15 Attempt to Clear Multifeed ...
Page 167: ...THEORY 6 17 2002 March Rev P 1908110 Figure 6 16 Monitor Feed Completion ...
Page 197: ...TROUBLESHOOTING 7 29 2004 December Rev S 1908110 THEORY ...
Page 300: ...SERVICE MANUAL 8 4 1908110 2004 December Rev S Figure 8 2 HOOD ASSEMBLY 2 3 4 5 6 1 ...
Page 310: ...SERVICE MANUAL 8 14 1908110 2004 December Rev S Figure 8 7 Accessory Parts 2 1 3 39 ...
Page 330: ...SERVICE MANUAL 10 2 1908110 2003 September Rev Q BLANK PAGE ...
Page 331: ......