134
Input of temperature-corrected value of calibration solution with simultaneous
display of cell constant (cell factor for toroidal sensors).
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_SOL
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Immerse sensor in cali-
bration solution. Enter the
temperature- corrected value
of the calibration solution
using the arrow keys (see
table).
Press enter to confirm.
Lower line: display of cell
constant (cell factor) and
temperature
Contacting conductivity
measurement (Cond)
The determined cell constant
is displayed.
The “hourglass” icon is
blinking.
Proceed by pressing enter
Inductive conductivity
measurement (CONDI)
The determined cell factor
and zero point are displayed.
The “hourglass” icon is
blinking.
Proceed by pressing enter
Calibration with Calibration Solution