103
Menu item
Action
Choices
Temperature
compensation
Select desired compensation
using
keys:
OFF: Temperature compensation
switched off
LIN: Linear temperature com-
pensation with entry of tempera-
ture coefficient and reference
temperature
nLF:
Temperature compensation for
natural waters to EN 27888
NaCl, HCL, NH3, NaOH:
Ultrapure water with traces of
impurity
(0 ... +120 °C / +32 ... +248 °F)
Temperature compensa-
tion of process medium
With linear compensation only:
Step 1:
Enter temperature compensation
of the process medium.
Enter
reference temperature
Step 2:
Enter reference temperature
Enter value using
keys
Press enter to confirm.
Permissible range 0 ... 199.9 °C
Temperature Compensation