Configuring a Cond Sensor
Overview of Cond Sensor Configuration
Cond sensor
Choices
Default
SNS:
2-ELECTRODE
4-ELECTRODE
MEMOSENS
2-ELECTRODE
2-EL / 4-EL
CELLFACTOR
00.0000 -
19.9999 c
01.0000 c
MEAS MODE
Cond
Conc %
Sal ‰
USP μS/cm
Cond
Cond
MEAS RANGE
x.xxx μS/cm
xx.xx μS/cm
xxx.x μS/cm
xxxx μS/cm
x.xxx mS/cm
xx.xx mS/cm
xxx.x mS/cm
x.xxx S/m
xx.xx S/m
xx.xx MΩ
xxx.x mS/cm
Conc
Solution
-01- (NaCl)
-02- (HCl)
-03- (NaOH)
-04- (H2SO4)
-05- (HNO3)
-06- (H2SO4)
-07- (HCl)
-08- (HNO3)
-09- (H2SO4)
-10- (NaOH)
-11-(H
2
SO
4
•SO
3
) (oleum)
-01- (NaCl)
TEMP UNIT
°C / °F
°C
TEMPERATURE
(EXT. only with I-input enabled via TAN)
AUTO
MAN
EXT (only if enabled via
TAN)
AUTO
AUTO
RTD TYPE
100 PT
1000 PT
8.55 NTC
30 NTC
Ni100
100 PT
MAN
TEMPERATURE
–50 ... 250 °C
(–58 ... 482 °F)
025.0 °C
(077.0 °F)