85
1) The cell constant can be modified by an entry in the configuration menu or
by calibration (one storage position). This means, a cell constant determined
by calibration is taken over by pressing enter during configuration. It remains
unchanged until a new value is entered.
2) For conductivity (μS/cm), the range selection determines the max. resolution.
If the selected range is exceeded, the device automatically switches to the next
higher range until the max. measurement limit is reached (9999 μS/cm).
This applies to display values and current outputs. The current outputs are
adjusted using a floating-point editor which allows settings over several decades.
The initial range of the editor is the selected range:
Selected
resolution
Displayed range (or floating-point editor)
Cond-Cond Configuration
3) Entries for calculating the consumption of the ion exchanger:
Activate with EXCHER CAP = ON. Messages in the Diagnostics / Monitor menu.
Configuration
Select
Default
COEFFICIENT
00.00 ... 99.99
FACTOR 1
FACTOR 2
FACTOR 3
0.001 ... 9.999
1.000
-C9-
ALCALISING
NaOH, NH3, LiOH
NaOH