Alarm
Menu item
Action
Choices
Alarm delay
Enter alarm delay using
keys.
Press enter to confirm.
Sensocheck
Select Sensocheck (continuous
monitoring of lines).
Select ON or OFF using
keys.
Press enter to confirm.
(At the same time, Sensoface is
activated. With OFF, Sensoface is
also switched off.)
Tempcheck
(see page 90)
To monitor the temperature
probe with TC OFF selected:
Select Tempcheck ON using
keys.
Press enter to confirm.
Now, the temperature probe will
be monitored.
Alarm Contact
The alarm contact is closed during normal operation (N/C).
It opens in the case of alarm or power outage. As a result, a failure
message is output even in the case of line breakage (fail-safe behav-
ior). For contact ratings, see Specifications.
Error messages can also be signaled by a 22-mA output current (see Error messages and Configuration
Output 1/Output 2).
Operating behavior of the alarm contact: see Operating States table.
The alarm delay time delays the color change of the display backlighting to red, the 22-mA signal (if
configured), and the alarm contact switching.
Configuring the Alarm