Specifications
Sensor
standardization
– Input of cell factor with simultaneous display of conductivity/temperature
– Input of conductivity of calibration solution with simultaneous display
of cell factor and temperature
– Product calibration
– Zero adjustment
– Installation factor
– Temp probe adjustment (10 K)
Permissible cell factor
00.100 ... 19.999/cm
Permissible transfer ratio
010.0 ... 199.9
Permissible zero offset
±0.5 mS/cm
Permissible installation
factor
0.100 … 5.000
Sensocheck
Monitoring of primary and secondary coils and lines for open circuit
and of primary coil and lines for short circuit
Delay
Approx. 30 s
Sensoface
Provides information on the sensor condition (zero point, cell factor, installation
factor, Sensocheck, TAG/GROUP sensor verification)
Sensor monitor
Direct display of measured values from sensor for validation
(resistance/temperature)
* User-defined