137
Product Calibration
Display
Action
Remark
The device returns to mea-
suring mode.
From the blinking CAL
mode indicator, you see
that product calibration
has not been terminated.
Product calibration step 2:
When the sample value has
been determined, open the
product calibration once
more
Display (3 sec)
Now the device is in
HOLD mode.
The stored value is displayed
(blinking) and can be over-
written with the lab value.
Press enter to proceed.
Display of new cell constant /
cell factor (based on 25°C).
Sensoface is active.
To exit calibration:
Select MEAS, then enter
To repeat calibration:
Select REPEAT, then
enter
End of calibration
After end of calibration,
the outputs remain in
HOLD mode for a short
time.