SIMDOS diaphragm metering pump
RC version – external drive
KNF Flodos BA_SIMDOS_EN_02_160044.docx
Translation of original operating manual
35
Digital input 2, Combined
The “Combined” setting activates the following functions on
application of a signal to digital input 2:
Interrupt a metering process
Reset of pump errors on signal edge
If the signal is applied for longer than 1 second, the pump will
operate in “Prime” at high speed for the duration of the signal.
Deletion of an active Master Stop (see section 7.7), pump is
now in “Pause” operation condition, and responds to external
control signals.
Fig. 54: Combined function sequence
START / STOP with footswitch via Level
4. Adjust Digital Input 2 „In_2“ to Level.
5. Connect the footswitch to the pump.
6. Press the „START“ button - pump changes status to „Pause“.
As long as the footswitch is pressed the pump will run.
When the Timer is switched on the pump will complete an
entire dosing cycle.
START / STOP with footswitch via Impulse
7. Adjust Digital Input 2 „In_2“ to Impulse.
8. Connect the footswitch to the pump.
9. Press the „START“ button - pump changes status to “Pause“.
Every time the footswitch is activated the pump turns on or off.
When the Timer is switched on the pump will complete an
entire dosing cycle.
Combined
Start/Stop level
Start/Stop pulse