SIMDOS diaphragm metering pump
Maintenance
KNF Flodos BA_SIMDOS_EN_02_160044.docx
Translation of original operating manual
37
9. Maintenance
9.1. Maintenance plan
Component
Maintenance interval
Pump
Regular inspections for external damage
or leaks.
Pump head
Clean if flow rate deteriorates, pump fails
to create vacuum or will not run.
Replace parts if necessary (see Chapter
11).
Pump diaphragm
-
Change the diaphragm after 1000
hours of use.
-
Change the diaphragm if the flow rate
changes or the pump leaks.
-
After changing the head parts several
times it is advisable to replace the
diaphragm.
Intake filter
(accessory)
Change if soiled.
Tab. 23: Maintenance plan
9.2. Cleaning
Flushing the pump
When pumping aggressive media, KNF recommends flushing
the pump with a neutral fluid under atmospheric conditions for
a few minutes before switching off, to extend the service life of
valves and diaphragm.
Cleaning the pump
Wipe the outside of the pump with a soft dry cloth. Ensure that
any cleaning solvents used will not damage the materials of
the pump head and crankcase housing (check material
resistance properties).
Cleaning the pump head
The pump head should be cleaned only if the pump is no
longer working properly (no suction, inadequate flow rate or
reverse suction of the pumped medium), or if the pump head
cannot be cleaned satisfactorily by flushing.
Remove pump head, clean and reinstall (see section 8.3).