SIMDOS diaphragm metering pump
RC version – external drive
KNF Flodos BA_SIMDOS_EN_02_160044.docx
Translation of original operating manual
33
Start/Stop level with timeout
1. Set digital input 1 “In_1” to Level.
2. Press “START” – the pump goes into “Pause” condition.
3. Apply signal 1. The pump will start to operate when the level is
connected to ground, and stop when the set time has expired.
If the pump time counter has been activated, the pump cannot
then be stopped with digital input 1. This input functions only
for starting the pump.
To stop the pump, use digital input 2, “Reset” (see section
7.20).
Fig. 51: Start ml/min with time counter level control
1. Set digital input 1 “In_1” to Pulse.
2. Press “START” – the pump goes into “Pause” condition.
3. Apply signal 1. The pump will start to operate when a pulse to
ground is generated, and stop when the set time has expired.
Fig. 52: Start ml/min with time counter pulse control
If the pump time counter has been activated, the pump cannot
then be stopped with digital input 1. This input functions only
for starting the pump.
To stop the pump, use digital input 2, “Reset”
(see section 7.20).
Start/Stop level with timeout
Start/Stop pulse with timeout