SIMDOS diaphragm metering pump
Operation
KNF Flodos BA_SIMDOS_EN_02_160044.docx
Translation of original operating manual
19
If it is possible that pressures in excess to the limits can occur
it is important to take the necessary measures, e.g. a pressure
relief valve or pressure sensor.
For further assistance contact your local KNF specialist.
Never apply positive pressure to the suction side of the pump.
A small positive pressure of less than 0.5 m water column will
cause the pump to block flow in the pumping direction. On the
application of any higher pressures than this, the pump will
allow the fluid to pass through in the pumping direction.
This can be prevented by using a suitable check / pressure
valve* on the outlet side of the pump.
* e.g. Pressure Control Valve FDV30 from KNF Flodos
7.2. Operating controls
1
STOP button
2
START button
3
Display
4
Control knob
Fig. 15: Operating controls
7.3. Main display
The main display shows the principal pump settings.
1
Specified flow
2
Flow unit
3
Time counter
4
Volumeter
5
External drive
6
Autostart
7
Analogue input
8
Digital input 1
9
Digital input 2
10
Operating condition
display
Fig. 16: Main display