Assembly and connection
SIMDOS diaphragm metering pump
KNF Flodos BA_SIMDOS_EN_02_160044.docx
14
Translation of original operating manual
6. Assembly and connection
All pumps must be installed only in accordance with the operating
parameters and conditions described under Technical data (see
Chapter 4).
Note the safety instructions in Chapter 3.
6.1. Assembly
Store at the assembly location prior to assembly so that the
pump is at the same temperature as the surrounding
environment.
Pump dimensions (see Fig. 10)
Fig. 10: Mounting measurements (measurement tolerances as per DIN
ISO 2768-1, tolerance class V)
Ensure that the installation location is dry, and protected from
water in the form of rain, spray, splashes and drips.
Select a level surface that will provide a firm base for the
pump.
When choosing the location, ensure that the hose connections
can be fitted without strain on the hoses. Do not pull on the
hoses, and avoid kinks and bends.
Protect the pump from dust.
Protect the pump from vibrations and impacts.
The pump is designed to be installed in an upright position.
Two additional mounting screws can be inserted from the
bottom of the pump base (see Fig. 10).
6.2. Electrical connection
Dimensions
Installation location
Installation position