SIMDOS diaphragm metering pump
RC version – external drive
KNF Flodos BA_SIMDOS_EN_02_160044.docx
Translation of original operating manual
31
Fig. 47: Analogue input
Starting and stopping the pump with an analogue signal
1. Set the analogue input to the correct signal type.
2. Press “START” – the pump goes into “Pause” mode.
3. Apply analogue signal. The pump now starts operating, and
goes into “Run” condition.
4. Reduce analogue signal to below off threshold – the pump
stops and goes into “Pause” condition.
After the pump has been stopped, there is a pause time of 0.2
seconds before it can be restarted.
Stopping the pump with the “STOP” button is called a “master
stop”. The pump can then only be restarted manually
(see section 7.7).
Master Stop is activated on the pump when supplied. For initial
start-up with external drive, it is therefore essential to press
“START” once to cancel the Master Stop condition.
For an external restart after a power cut, use Autostart (see
section 7.18) or Reset on digital input 2 (see section 8.3).
START/STOP analogue signal
Master Stop