Rear panel I/O
AZURA® Detector ECD 2.1 Instructions, V6685
Fig. 61
Dialogue Elite Events tab with an example of two programmed
events switching the valve to Inject (t = 0.01 min) and back to Load
(t = 0.05 min).
For more detailed information about valve control of the elec-
trically-actuated VICI valves in the ClarityChrom® chromatogra-
phy software or Dialogue Elite please refer to the corresponding
instructions.
11.4 ANALOG DATA connector
The ECD 2.1 is equipped with an analog data output connector to provi-
de the measured signals in millivolts (mV) for users who work stand-alone
without PC control with the instrument. The ANALOG DATA connector
(D9-female) can either be connected to an X-Y recorder or A/D converter.
The ANALOG DATA output can supply either a non-manipulated signal
directly from the I/E converter, or the processed data signal by the CPU of
the ECD 2.1. The type of output can be selected from the CONFIG screen
by setting the parameter Vout source to either DAC or I/E.