Troubleshooting
AZURA® Detector ECD 2.1 Instructions, V6685
Fig. 76
Disconnect tubing connection
Disconnect the outlet tubing from the flow cell.
Connect the other end of the tubing (of the tubing connected to the
inlet of the flow cell) to the outlet of the flow cell.
The fluidic path of the flow cell is now completely isolated from the
rest of the LC system.
Record a run (without injection of sample) to measure/evaluate the
background cell current (Icell) and noise.
Compare the obtained values of the cell current and noise with the
values observed before the stop flow with the pump on.
Results:
Icell: In case a significant drop in Icell is observed, for example a drop of
more than 50%, it is an indication that the problem is not flow cell related
but originates from other parts of the LC system. The most obvious rea-
sons for a high background current are electrochemically active contami-
nants in the mobile phase, column bleeding, leaking pulse damper. This
causes can be systematically eliminated by replacing the mobile phase or
disconnecting column or pulse damper etcetera, and reevaluate the cell
current.
Fig. 77
Significant drop in Icell