72
Rear panel I/O
AZURA® Detector ECD 2.1 Instructions, V6685
Fig. 59
Example of the valve flow path of a 6-port valve.
In Dialogue Elite the valve status/position is shown under the monitor tab.
To make it visible in the left table, check the valve position checkbox in
the item selector (by default the valve position is not shown).
Fig. 60
Dialogue Elite Monitor tab with valve status (position) displayed.
An electrically-actuated VICI valve can also be controlled via Dialogue
Elite under the detector tab, see Fig. 56. Note that the valve will switch
immediately after a change from Load to Inject (checkboxes) or vice ver-
sa, no ‘Send to device’ action is required.
An electrically-actuated VICI valve can also be controlled via timed events
under the events tab in Dialogue Elite. In the Event editor the event ‘valve
position’ is available to switch the valve in the Load/inject position. See