Do not place object/instruments on top of the detector. Objects can be
placed on any side of the detector; however, make sure these objects are
placed at a distance of 5 cm from the detector, in case that objects are
placed at only one side of the instrument. and 10 cm from the instrument,
if objects are placed on both sides of the instrument.
3.3 Mains connection
Check that fuses and voltage range on the rear side of the instrument
match that for the power outlet to be used.
This device complies with the requirements of EN 55011 group 1
class A ISM equipment. In a domestic environment this product may
cause radio interference in which case the user may be required to
take adequate measures on his own expenses. Understanding power
surges. Power surges, line spikes, and transient energy sources can
impede instrument operation. Ensure that the instrument's electrical
supply is adequately protected from these conditions and properly
grounded. If necessary connect the instrument to a filtered mains
inlet.
Power supply and protective earth: The system must be connected
to a suitable mains power supply with a correctly installed protective
earth conductor. Never use the system without a properly connected
protective earth conductor.
Leave the instrument powered off until specifically mentioned in the pro-
cedure below.
3.4 PC connection
Follow the instructions in this paragraph when the instrument is used with
PC control over LAN using the Dialogue Elite or ClarityChrom® Chroma-
tography Software. This section can be skipped if the instrument is used
stand-alone. To be able to communicate over LAN a computer is required
with a free (PCI, PCI Express or PCI-X) LAN port.
The ECD 2.1 has a fixed IP address: 192.168.5.1, with subnet mask:
255.255.255.0. Gateway and DNS are not filled in.
The instrument is standard delivered with a special crossover LAN (UTP)
cable which is part of the ECD 2.1 accessory kit that is shipped with the
detector.
To insure stable and error-free communication only use the manufac-
turer-supplied LAN connectivity kit to connect the ECD 2.1 to LAN.
Create a small dedicated local area network to connect the ECD 2.1
to the PC. Do not connect the ECD 2.1 over a company Local Area
Network. If needed a second network adapter with a different (uni-
que) IP address range can be applied.
12
Installation
AZURA® Detector ECD 2.1 Instructions, V6685
Summary of Contents for Azura ECD 2.1
Page 1: ...HPLC Detector ECD 2 1 Instructions Document No V6685 ...
Page 32: ...28 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Pulse mode ...
Page 33: ...29 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Scan mode CONFIC menu ...
Page 34: ...30 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 DIAG menu ...