Troubleshooting
AZURA® Detector ECD 2.1 Instructions, V6685
Baseline oscillations
Possible cause
Remedy
Malfunctioning pump
(regular pattern)
Check pump (seals, valves)
Over-tightened cell
bolts
Adjust cell bolts, check pump pressure
Air bubbles in cell or
REF
Maintenance REF
Temperature
oscillations
Set oven temperature
Contaminated buffer
(high I
cell
)
Replace buffer, do not recycle the buffer
Fouled WE
Clean WE
Fe++ in buffer
Add EDTA, passivate metal parts with HNO
3