The ECD 2.1 must only be used with appliances and power sources with
proper protective grounding to prevent damage through build-up of sta-
tic electricity. The power source should exhibit minimal power transients
and fluctuations. If necessary connect to a filtered mains socket.
Replace blown fuses with fuses of proper type and rating as indicated on
the rear panel and as listed in the list of accessories and spares (appendix
D). The fuse holder is integrated in the mains connector. Ensure that the
instrument is never put in operation with fuses of a different type. This
could cause fire. Only use manufacturer-supplied I/O cable(s) to connect
with other devices. Thoroughly connect shielding to common. Manufac-
turer will not accept any liability for damage, direct or indirect, caused by
connecting this instrument to devices and with cables which do not meet
relevant safety standards.
Solvents
The solvents used may be flammable, toxic or corrosive. The room in
which the system is installed should be well ventilated to prevent that
solvent vapors cause poisoning or ignite and cause a fire. Use of open fire
in the vicinity of this system must be strictly prohibited. Do not install the
system in the same room with any other equipment that emits or could
potentially emit sparks. Provide protective equipment near the instru-
ment, when solvent gets into the eyes or on the skin, it must be flushed
away immediately. Provide equipment, such eye wash stations and safety
showers, as close to system as possible. Use proper eye and skin protecti-
on when working with solvents. Additional safety requirements or protec-
tion may be necessary depending on the chemicals used in combination
with this equipment. Make sure that you understand the hazards associa-
ted with the chemicals used and take appropriate measures with regards
to safety and protection.
Sample containers (vials) should be sealed to minimize any risks related
to solvent vapor.
5
General information
AZURA® Detector ECD 2.1 Instructions, V6685
Summary of Contents for Azura ECD 2.1
Page 1: ...HPLC Detector ECD 2 1 Instructions Document No V6685 ...
Page 32: ...28 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Pulse mode ...
Page 33: ...29 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Scan mode CONFIC menu ...
Page 34: ...30 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 DIAG menu ...