General Technical Data
Designation
All suspension arm systems
Operating mode
suitable for continuous operation
Rating plate
The rating plate is located on the upper extension
arm, see section 4.1 “Data on the Rating Plate”,
page 25.
Max. load-bearing capacity of the various spring
versions
30 – 60 kg
70 – 110 kg
120 – 180 kg
50 – 80 kg
80 – 135 kg
Electrical data
Nominal voltage
100–240 VAC
Nominal frequency
60/50 Hz
Rated output
max. 220 W
Indirect extension arm lighting
2–4 light panels
(operating voltage 12 VDC, each with 2 light
panels connected in series to 24 VDC)
12 VDC
Sound pressure level
A level of 65 db(A) (EN ISO 3746) is not exceeded
Manual forces (when the electromagnetic brake
is released depending on position and payload)
3.5–40 Nm
Protection class acc. to IEC 60601-1
I
IP classification as per IEC 60529
IP 20
Approvals of the standard design
Recognized NRTL component
Ambient conditions
Storage and transportation
-25°C to +70°C (-13°F to +158°F) at 10–75%
humidity (relative) and air pressure 500–1,060 hPa
Operation
+10°C to +40°C (+50°F to +104°F) at 30–75%
humidity (relative) and air pressure 700–1,060 hPa
(corresponds to a maximum operating altitude of
3,000 m)
Conformity to standards
see section 3.5 “Standards and Guidelines”, page 23
Medical Device Directive 93/42/EWG
Classification I
Permissible adaptions