Via the RaZON
+
web interface under System configuration/communication, the communication parameters can be changed. A
complete description of the serial communication protocol can be found in the Appendices.
Pin
Color
Signal
Direction
Description
Host connection
1
Red
2
Blue
3
Green
4
Yellow
RD+
input
RS-485 Receive data (B) RS-485 Transmit Data (Z)
5
Grey
RD-
input
RS-485 Receive Data (A) RS-485 Transmit Data (Y)
6
Brown
SG
RS-485 signal ground
RS-485 signal ground
7
White
TD+
output
RS-485 Transmit Data (Z) RS-485 Receive data (B)
8
Black
TD-
output
RS-485 Transmit Data (Y) RS-485 Receive Data (A)
Pin description of the RS-485 connector to HOST interface
The indicated colors above apply to the standard yellow 8 wire cable as used for the SMP pyranometers (with 8 pin Binder 712
connector). To use this port as 2 wire RS-485, connect the RD+ and TD+ together (as D+) and the RD- and TD- together (as D-) and
select in the webpage under system configuration/ Communication / Transmission, the 2-wire option. This switches local echo off.
6.3 Serial interface, AUX in port
The AUX IN port is intended for connection of sensors or weather stations that use the Modbus® RTU protocol. In the first release of
the RaZON
+
, this feature is not yet implemented.
1
Red
2
7
8
5
4
6
3
Blue
White
Black
Grey
Yellow
Brown
Green
Pin Layout (instrument)
Summary of Contents for RaZON+
Page 1: ...RaZON Solar Monitoring System Instruction Manual ...
Page 2: ...2 ...
Page 4: ...4 ...
Page 6: ...6 ...
Page 12: ...12 ...
Page 42: ...42 ...
Page 44: ...44 ...
Page 50: ...50 ...
Page 52: ...52 ...
Page 71: ...71 Example of a logged ASCII data file via Wi Fi Logging Eport Log File ...