3.13 Connecting DC power cable
This section will explain how to make and connect your power cable to power the RaZON
+
.
RaZON
+
is supplied with a waterproof connector for 24 VDC power. This connector is in accordance with safety regulations for outdoor
usage. The voltage drop over the cable should not cause the supplied voltage to be out of the operating limits (VDC should be between
20 to 30 VDC). The 24 VDC power connector is male (chassis part is female). The DC power connector is suitable for cable diameters
between 6 and 12 mm. For 6 to 9 mm and 9 to 12 mm different sealing rings are used. The DC power connector is a male type
Hirschmann CA 3 LS connector. The maximal cable resistance should be less than 2 x 0.5 Ω.
Please make sure the used cable is flexible enough to withstand the daily movement of the tracker. The number of strands per wire is
specified on the next page.
Pyrheliometer connection
Pyranometer connection
RS-232 / RS-485 Aux in
M4 bolts
Cable mounting
East mark on flange (E)
RS-485 to host
Reset switch
Ethernet 10/100
DC Power input 20 to 30V, 15W
East
USB WiFi connector