Verification of Autotest range, tone detector and linearity test
Test below is to verify Autotest range, tone detector operation and
linearity for Power Meter. Connect the Power Meter to Attenuator to
Light Source
Power Meter Attenuator Light Source
2.
Switch on all instruments.
3.
Set the Light Source to 1310nm and allow to warm up. Press
[Autotest].
4.
On the attenuator, set the wavelength to 1310nm and
attenuation to 5 dB.
5.
Make sure that Autotest is updating. Note the measured optical
power at 1310 nm.
6.
For single mode instruments, repeat the power readings for
attenuator settings of 15 dB, 25 dB and 35 dB. For instruments
with H series meter, repeat the readings for an attenuator
setting of 15 dB only.
7.
For multimode instruments, repeat the power readings for
attenuator settings of 15 dB, 20 dB.
8.
Repeat the above for 1550nm as required, however, this test
typically only needs performing at one wavelength.