CALIBRATION AND MAINTENANCE
KI2000 UM-6
-
51 -
There are no internal user adjustments. Calibration is performed
without opening the instrument.
Before commencing calibration:
Clean all optical connectors very carefully.
Ensure that all devices have been at a stable room
temperature for over an hour, and that the light source is
fully warmed up at the wavelength to be calibrated.
Note: instrument has to be powered externally while doing manual
calibration.
Perform calibration as per instructions below. When calibration is
completed, place an anti-tamper label over the hole. Do not forget to
update your calibration records and to schedule the next calibration
service.
Power Meter Calibration
Calibration is a transfer process. It is performed by setting up a light
source at a stable, but non-critical power level between 0 and -30 dBm,
and adjusting the meter reading to the same value as that shown by a
reference meter.
Required are laser and LED light sources with accurate, calibrated
wavelengths and good power stability, a power meter with appropriate
calibrated wavelengths, single mode and multimode test leads. Check
the calibration certificates on your reference equipment to ensure
current validity.
Known calibration constants can be re-entered directly without using
other equipment. This is useful in case old calibration constants are to
be put back.
Record the existing calibration offsets, re-enter or adjust known
offsets at this point or calibrate the meter at the selected wavelength
as follows:
Remove batteries.
Using tweezers, place a link across gold plated calibration
contacts inside battery compartment, Figure 2. Turn on
external power via micro USB.
Figure 2. Link has to be placed across gold plated calibration contacts
Press [-] or [+] to set the wavelength to be calibrated.
Record the light source power measured by reference
power meter.
Transfer this power level to the meter to be calibrated.
To modify calibration, press [F3] or [F4].
Gold plated
calibration
contacts
Summary of Contents for KI 2300 Series
Page 6: ...KI2000 UM 6 4 KI2400 2800 HAND HELD FIBER SOURCE ...
Page 7: ...KI2000 UM 6 5 KI2300 2700 HAND HELD LOSS TEST METER ...
Page 39: ...ACCURACY CONSIDERATIONS KI2000 UM 6 37 ...
Page 63: ...PERFORMANCE VERIFICATION TESTS KI2000 UM 6 61 Table 2 General Test Record ...
Page 72: ...NOTES KI2000 UM 6 70 Please use these pages to write notes ...
Page 73: ...NOTES KI2000 UM 6 71 ...
Page 74: ...NOTES KI2000 UM 6 72 ...