ACCURACY CONSIDERATIONS
KI2000 UM-6
-
36 -
DWDM Loss and Power Measurements
The DWDM bands are typically within 1525 ~ 1610 nm. For this
application, an InGaAs detector calibrated at 1550 nm gives good
absolute accuracy at all DWDM wavelengths. Testing optical loss at
1610 or 1625 nm is common, since this represents the worst case
attenuation in most systems.
High power InGaAs detectors may be required to measure the system
power level.
Ge detectors should not be used for accurate testing on WDM systems
beyond 1550 nm, since inaccuracies become large and unpredictable.
CWDM Loss and Power Measurements
The CWDM band is from 1270 ~ 1610 nm at 20 nm spacing, however,
1490 ~ 1610 nm is more common since it avoids the water absorption
peak at 1383 nm.
The most common operating wavelengths for FTTH PON systems are
currently 1310 / 1490 / 1550 nm.
As in DWDM application, an InGaAs detector gives good absolute
accuracy.
Because of the broad operating wavelengths encountered, the power
meter calibration wavelength needs to be matched to the measured
wavelength, or absolute errors may be excessive.
Many systems are measured at 1390 nm to test for the water peak,
and at 1625 nm to measure the high end attenuation limit.
At 1625 nm, the fiber attenuation is actually very sensitive to
wavelength, since the silica infra-red absorption zone is reached. So
the wavelength tolerance of this particular laser is a major issue,
especially since many older source instruments have a tolerance of 30
nm.
Summary of Contents for KI 2300 Series
Page 6: ...KI2000 UM 6 4 KI2400 2800 HAND HELD FIBER SOURCE ...
Page 7: ...KI2000 UM 6 5 KI2300 2700 HAND HELD LOSS TEST METER ...
Page 39: ...ACCURACY CONSIDERATIONS KI2000 UM 6 37 ...
Page 63: ...PERFORMANCE VERIFICATION TESTS KI2000 UM 6 61 Table 2 General Test Record ...
Page 72: ...NOTES KI2000 UM 6 70 Please use these pages to write notes ...
Page 73: ...NOTES KI2000 UM 6 71 ...
Page 74: ...NOTES KI2000 UM 6 72 ...