PERFORMANCE VERIFICATION TESTS
KI2000 UM-6
-
56 -
Optical Light Source
Output Power (CW) Test
1.
Connect the Light Source to the Power Meter as shown.
2.
Switch on the instruments.
3.
Set the Optical Power Meter to 1310 nm
4.
On Light Source, enable the source and set the wavelength to
1310 nm.
5.
Note the measured power level value in the test report in
Table 5.
6.
Repeat the above on 1550 nm.
Light source
Power Meter