7-6 Maintenance
OT01-PCR-LA/2
Symptom3: A voltage different from the set voltage is generated
(twice or 1/2).
Symptom4: The voltage display becomes abnormal in the line
voltage display mode.
Check item
Possible cause
Remedy
Location and condition
of the item concerned
Check results
Different voltage display
mode.
No
The phase voltage
display mode sets the
phase voltage, while
the line voltage dis-
play mode sets the
line voltage.
Select the correct voltage display
mode. See “6.2 Switching of Out-
put Voltage Display”.
Check item
Possible cause
Remedy
Location and condition
of the item concerned
Check results
Check whether the
SENSING terminal
boards are properly con-
nected.
No
Improper cable con-
nection to the SENS-
ING terminal board,
which disables line
voltage measurement
Connect the signal connection
cables to the SENSING terminal
boards properly. See “2.5.3 Signal
Connection Cables”.