KJM6335
Operation
3-9
When inputting RF output after the pick-up, always use the
instrument with SYMMETRY mode set to AUTO or OFFSET.
When SYMMETRY mode is set to AUTO, the slice level of the
instrument will automatically follow up the symmetry level of the
RF signal.
When SYMMETRY is set to OFFSET, the slice level will
automatically follow up the symmetry level of the RF signal, but
you may offset the automatic follow-up level by using the
SYMMETRY LEVEL variable resistor.
For detail, see 3.7 "Symmetry and Slice Level" and 2.1 " Front
Panel".
For setting the delay when DELAY mode is set to MANUAL, see
3.8 "Adjusting Delay".
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
NOTE
• The characteristics of the meter comply with Compact
Disc Reference Measuring Methods Specification
Guideline Ver.1.0 May 1999. A difference between the
characteristics of the symmetry follow-up circuit (slicer
circuit) in the CD player and those of the instrument will
result in a difference in measured values with respect to
3.4.1 "Measuring to Use RF Signal after Slicing and
Regenerated Clock Signal".
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○