KJM6335
Names and Functions of Controls
2-9
13 MEMORY
Allows you to store in memory up to four types of instrument panel
settings.
• MEMORY SELECT key
Selects a setup memory address. When this key is pressed, the
setup memory address changes to 1, 2, 3, and 4 in sequence.
• MEMORY LEDs
A selected address is indicated by the relevant LED at the left of
MEMORY SELECT key. The LED blinks for three seconds after
selection of the address and will change from blinking to
continuously lit if you recall or store the panel settings during this
period.
If recall or store is not performed within three seconds after the
selection, the LED returns to the state it was in before the
MEMORY SELECT key was pressed.
Further, if any operation other than recall or store is carried out
while the LED is blinking, the LED goes off.
• MEMORY RECALL / STORE key
Used to recall the memory contents.
Pressing this key with the 15 SHIFT key held down causes the
panel settings to be stored at a selected setup memory address.
14 GPIB
• GPIB LOCAL key
Pressing this key while the instrument is in the remote state
controlled by GPIB (the RMT LED at the left of the key is lit)
causes the instrument to return to control from the front panel
(local status).
• RMT LED
Lights up when the instrument is in the remote state controlled
using the GPIB interface.
If an invalid operation is performed (such as operating the panel in
the key lock state), this LED blinks regardless of the GPIB
condition.