2-8
Names and Functions of Controls
KJM6335
12
JUDGE
13
MEMORY
14
GPIB
17
CLOCK
16
RF SIGNAL
15
SHIFT
RF
CLK
INHIBIT
12 JUDGE
• JUDGE LEVEL key
The judgement level is indicated on the meter while this key is
pressed. The meter's scale is set with the 10 METER SCALE key.
• JUDGE LEDs
GO LED lights up when a measured value is lower than a set value
of judge level; NO GO LED(red) lights up when the measured
value is larger than the set value.
If no signal is input or the clock frequency is outside the specified
range, the NO GO LED blinks.
• JUDGE LEVEL variable resistor
This variable resistor is used to set a judgement level.
The judgement level can be individually set in the unit (% or ns)
which is set with the 10 METER SCALE key.
Even when you change the unit, the set value of each judgement
level is retained.