STAT:QUES:COND
Queries the condition of the QUEStionable status register.
Command
STATus:QUEStionable:CONDition?
Response
Returns the condition of the QUEStionable status register in the <NR1> form in
response to STAT:QUES:COND?.
STAT:QUES:COND:LIST:OVP
STAT:QUES:COND:LIST:UVP
Queries the condition of bit 9 (UVP) or bit 10 (OVP) of the QUEStionable status
register.
Command
STATus:QUEStionable:CONDition:LIST:OVP?
STATus:QUEStionable:CONDition:LIST:UVP?
Response
Returns the current condition in the <String> form in response to
STAT:QUES:COND:LIST:OVP?.
Returns the current condition in the <String> form in response to
STAT:QUES:COND:LIST:UVP?.
(Example)
If OVP is activated on CH31, CH122, and CH205, and UVP is not activated,
the following is returned in response to
STAT:QUES:COND:LIST:OVP?;UVP?.
"31,122,205";"".
STAT:QUES:ENAB
Sets the enable register of the QUEStionable status register.
Command
STATus:QUEStionable:ENABle <NR1>
STATus:QUEStionable:ENABle?
Parameter
V
al
u
e:
0 to 32 767
Response
Returns the enable register of the QUEStionable status register in the <NR1> form
in response to STAT:QUES:ENAB?.
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......