KFM2151
4-19
Remote Control
VOLT:PROT:LOW
Sets the voltage protection (UVP) of the channel specified by CHAN:VOLT:NSEL.
Command
[SENSe:]VOLTage[:DC]:PROTection:LOWer
{<numeric>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:PROTection:LOWer? [{MIN|MAX}]
Parameter
V
al
u
e:
-2 to 200
The default value is -2.
Unit:
V
Table 4-3 shows the settings when *RST, MEAS?, CONF:VOLT, or CONF:VOLT:ALL is
sent.
Response
Returns the voltage protection (UVP) in the <NR3> form in response to
VOLT:PROT?.
4.5.3
Releasing the Protection Function
INP:PROT:CLE
Releases the alarm.
Command
INPut:PROTection:CLEar