4-1
8
KFM2151
Remote Control
VOLT:RANG:AUTO
Sets the voltage range of the channel specified by CHAN:VOLT:NSEL to auto.
Command
[SENSe:]VOLTage[:DC]:RANGe:AUTO {ON|OFF|1|0}
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
Parameter
V
al
u
e:
ON(1)
Auto range on (default)
OFF(0)
Auto range off
Table 4-3 shows the settings when *RST, MEAS?, CONF:VOLT, or CONF:VOLT:ALL is
sent.
Response
Returns the auto range setting in the <NR1> form in response to VOLT:RANG:AUTO?.
VOLT:PROT
Sets the voltage protection (OVP) of the channel specified by CHAN:VOLT:NSEL.
Command
[SENSe:]VOLTage[:DC]:PROTection[:UPPer]
{<numeric>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:PROTection[:UPPer]? [{MIN|MAX}]
Parameter
V
al
u
e:
-2 to 200
The default value is 200.
Unit:
V
Table 4-3 shows the settings when *RST, MEAS?, CONF:VOLT, or CONF:VOLT:ALL is
sent.
Response
Returns the voltage protection (OVP) in the <NR3> form in response to
VOLT:PROT?.