KFM2151
4-17
Remote Control
CHAN:ROUT:NSEL
Specifies the logical channel for impedance scanning. However, the logical channel
cannot be specified, if the specified channel is not a scanned source.
Command
CHANnel:ROUTe:NSELect <chan>
CHANnel:ROUTe:NSELect?
Parameter
<chan>
The default value is 1.
If the specified channel is not a scanned source, an SCPI error (-
221, "Settings conflict") is returned.
The settings after sending *RST are indicated in Table 4-3.
Response
Returns the logical channel of the specified impedance scanning in the <NR1> form
in response to CHAN:ROUT:NSEL?.
VOLT:RANG
Sets the voltage range of the channel specified by CHAN:VOLT:NSEL.
Command
[SENSe:]VOLTage[:DC]:RANGe[:UPPer]
{<numeric>|MIN|MAX|DEF}
[SENSe:]VOLTage[:DC]:RANGe[:UPPer]? [{MIN|MAX}]
Parameter
V
al
u
e:
2, 20, or 200The default value is 200.
Table 4-3 shows the settings when *RST, MEAS?, CONF:VOLT, or CONF:VOLT:ALL is
sent.
Response
Returns the present range setting in the <NR1> form in response to VOLT:RANG?.