KFM2151
4-43
Remote Control
4.9
Tutorial
This section describes the procedures to make voltage measurements through
remote control on the KFM2151.
4.9.1
Simple measurement procedure (MEASure)
For simple measurement procedures, use the MEASure:VOTLage? query. When
you use the MEASure:VOLTage? query, all measurement conditions except the
range setting are set to default, and the measured data can be retrieved easily.
Example:
CHAN:VOLT:NSEL 1 ' Select logical channel 1.
MEAS:VOLT?
' Start voltage scanning on all
activated channels, and query the
measured data of the selected
channel.
Response example: 1.72000E00
Example 2:
MEAS:ARR:VOLT?
' Start voltage scanning on all
activated channels, and query the
measured data of all activated
channels.
Response example: 1.72000E00,1.69000E00,2.92000E00
Because the MEASure:VOLT? query inherently calls the CONFigure command, the
voltage measurement range is set as shown in Table 4-20.
Ta
b
le 4-20
V
oltage meas
u
rement range set
b
y MEAS
The MEAS:VOLT? and CONF commands have parameters (can be omitted).
MEASure:VOLT? [<numeric>]
CONFigure:VOLT [<numeric>]
Specify the estimated range of the measured data in <numeric>.
If <numeric> is set to DEFault or is omitted, AUTO range is used.
Setting
command
V
al
u
e
V
oltage meas
u
rement range
V
OLT:RA
N
G
Specified
v
al
u
e (AUTO
w
hen omitted)
V
oltage protection (O
V
P)
V
OLT:PROT
200
V
oltage protection (U
V
P)
V
OLT:PROT:LO
W
-2
Summary of Contents for KFM2151
Page 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Page 4: ...ii KFM2151...
Page 14: ...xii KFM2151 Contents...
Page 35: ...Operation This chapter mainly describes various setup procedures...
Page 54: ...3 20 KFM2151 Operation...
Page 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Page 100: ...4 46 KFM2151 Remote Control...
Page 111: ......