Operation and Features
2
Keysight U8030A Series User’s Guide
59
System-Related Operations
Reset to the factory defaults
To reset to factory defaults, press and hold the
[All On/Off]
button when powering
on the unit until the display shows
dOnE
.
– The OVP and OCP are disabled, and their trip levels are set to maximum (i.e.,
10% above the CV and CC rated values).
– All memory storage locations are cleared.
– The voltage limit value is set to 0, while the current limit value is set to the
maximum rated value.
– The calibration data are not affected by the factory default reset.
Power-on self-test
A power-on self-test occurs automatically when the power supply unit is turned
on. This test ensures the unit is operational. The power-on self-test checks for
flash data memory and checks the voltage and current outputs when the unit is
turned off.
Display the firmware and board versions
To display the firmware and board versions, press and hold the
[Over Voltage]
and
[Display Limit]
buttons simultaneously when powering on the unit.
– The display shows the front panel firmware version on line 1 for one second.
– After that, the display shows:
– the firmware and board versions of analog board 1 on line 1, and
– the firmware and board versions of analog board 2 on line 2 for one second.
NOTE
If the test fails, the display will show
Err
along with the error code. For more
information, refer to the
Summary of Contents for U8030A Series
Page 2: ...Keysight U8030A Series Triple Output DC Power Supply User s Guide ...
Page 9: ...8 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 13: ...12 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 15: ...14 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 17: ...16 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...