Constant Current Operation
The following steps show you how to perform the constant current (CC) operation.
1 Short the binding posts.
– Press the
[POWER]
button to turn off the power supply.
– Connect a short circuit between the positive (+) and negative (—) binding
posts of your desired output.
2 Turn on the power supply.
– Press the
[POWER]
button to turn on the power supply.
– The power supply will then perform a self-test (self-test is not indicated at
the display).
– The display turns on all annunciators briefly.
– The outputs are disabled by default (the
OFF
annunciator turns on).
– The display shows
OFF
for both outputs (line 1 and 2).
3 Select an output channel.
– Press
[1]
to select Output 1, or
– Press
[2]
to select Output 2.
– The corresponding annunciator (
OUT1
or
OUT2
) turns on.
4 Set the display to the limit mode.
– Press the
[Display Limit]
button to set the display to the limit mode. The
LIMIT
annunciator turns on.
– The display will show the voltage and current limit values for the selected
output.
– The display for the non-selected output remains unchanged.
Summary of Contents for U8030A Series
Page 2: ...Keysight U8030A Series Triple Output DC Power Supply User s Guide ...
Page 9: ...8 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 13: ...12 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 15: ...14 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 17: ...16 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...