Operation and Features
2
Keysight U8030A Series User’s Guide
55
Programming the Overcurrent Protection
Overcurrent protection guards the load against output currents reaching values
greater than the programmed protection level.
The following steps shows you how to enable and disable the overcurrent
protection (OCP), how to set the OCP trip level, and how to clear the overcurrent
condition.
Set the OCP trip level and enable the OCP
1 Select an output channel.
– Press
[1]
to select Output 1, or
– Press
[2]
to select Output 2.
– The corresponding annunciator (
OUT1
or
OUT2
) turns on.
2 Set the OCP trip level.
– Press the
[Over Current]
button.
– The
LIMIT
annunciator turns off.
– The corresponding OCP annunciator (
OCP1
or
OCP2
) will blink
continuously.
– The display will show the OCP value for the selected output.
– The
A
annunciator will blink continuously.
3 Turn the knob to ad just the OCP value.
– While the
A
annunciator is blinking, turn the knob and adjust to the desired
OCP value.