Introduction
1
Keysight U8030A Series User’s Guide
21
3 Verify that the correct power-line fuse is installed.
The correct fuse for your country is installed when the power supply is shipped
from the factory. See the table below to replace the fuse for your power supply.
Checking the instrument output
The following procedures check to ensure that the power supply develops its
rated outputs and properly responds to operation from the front panel. For
complete performance and verification tests, refer to the
U8030A Series Service
Guide
.
From the front-panel VFD (vacuum-fluorescent display), you can monitor actual
values of output voltage and current (meter mode) or voltage and current limit
values (limit mode).
Voltage output check
The following steps verify basic voltage functions with no load.
1
Turn on the power supply.
The power supply starts up for the first time in the power-on (reset) state — all
the outputs are disabled (the
OFF
annunciator is turned on).
2
Enable the outputs.
Press the
[All On/Off]
button once. Notice that the display is in the meter
mode.
Table 1-1
List of rated fuse for line voltages
Model
Keysight part number
Part description
U8031A/
U8032A
2110-1504
FUSE 1.0 A TIME-DELAY 0.0757
Ω
20 mm
×
5.2 mm
×
5.2 mm
NOTE
If an error is detected during the output checkout procedure, the
Err
annunciator
will turn on. For more information, refer to the
Summary of Contents for U8030A Series
Page 2: ...Keysight U8030A Series Triple Output DC Power Supply User s Guide ...
Page 9: ...8 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 13: ...12 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 15: ...14 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 17: ...16 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...